Sinuo Testing Equipment Co. , Limited

Sinuo Testing Equipment Co., Limited

Manufacturer from China
Verified Supplier
6 Years
Home / Products / Switch Life Tester /

IEC 61032 Figure 13 Test Probe 19 Ф5.6mm Children Finger Probe

Contact Now
Sinuo Testing Equipment Co. , Limited
Visit Website
City:guangzhou
Province/State:guangdong
Country/Region:china
Contact Person:MsZoe Zou
Contact Now

IEC 61032 Figure 13 Test Probe 19 Ф5.6mm Children Finger Probe

Ask Latest Price
Video Channel
Brand Name :Sinuo
Model Number :SN2210-13
Certification :Calibration Certificate (Cost Additional)
Place of Origin :China
MOQ :1
Price :Negotiable
Payment Terms :L/C, T/T, Western Union, MoneyGram
Supply Ability :100 sets per month
Delivery Time :3 Days
Packaging Details :Aluminium Box
Applied Standard :IEC 61032
Applied Figure :Figure 19
Finger Probe Diameter :Ф5.6mm
Finger Probe Length :44mm
Probe Material :Stainless steel
Handle Material :Nylon
more
Contact Now

Add to Cart

Find Similar Videos
View Product Description

IEC 61032 Figure 13 Test Probe 19

Product Overview:

This probe conforms to the standard requirements of IEC 61032 figure 13 test probe 19, IEC 60065:2014 clause 9.1.1.3, etc. This probe is intended to simulate access to hazardous parts by children of 36 months or less.

Probe Dimension:

Hemispherical radius (mm) R2.8±0.05
Finger probe diameter (mm) Ф5.6±0.1
Finger probe length (mm) 44±0.15
Axes of joints distance (mm) 14.67
Collar diameter (mm) Ф25.9±0.3
Collar length (mm) 101.6±0.3
Extension handle diameter (mm) Ф25.4±0.3
Extension handle length (mm) 464.3±0.8
Finger probe material (mm) Stainless steel
Collar and handle material (mm) Nylon
Operation:

Please refer to the relevant standards.

Take IEC 60065:2014 clause 9.1.1.3 as an example:

The test is repeated using small finger probes according to test probes 18 and 19 of IEC 61032:1997. This does not apply if the intended conditions of use prevent the apparatus from being accessed by children.

For verifying whether the LIVE PARTS touched, the probe is equipped with amphenol connector, recommended to work with our anti-shock probe experiment device or a 40~50V voltage indicator. Showing as following reference picture: