IEC60529 IPX1~2 Vertical Rain Drip Test System To Verify Water Drops
Product Information For IPX1 IIPX1~2 Vertical Rain Drip Test System:
This test device is to verify Protection against Vertically Falling Water Drops. It is designed and manufactured in accordance with the standard requirement of IEC60529 IPX1 and IPX2, IEC60335-1 clause 15.etc.
The rise and fall of the drip box is controlled by the motor. The electrical control cabinet is with PLC and touch screen.
1. Stainless steel fine filter for dirt filtration.
2. Adopts compressed air (compressed air should be prepared by user) to blow out the water in the needles, to prevent water reactive substances and atmospheric blocking needles.
3. Stainless steel general medical needle nozzles, easy to buy, not easy to rust.
4. Needle nozzles and needle nozzle seats using taper connection, can be easily replaced without the tool, individual damage to the needle nozzle can be easily replaced quickly.
Technical Parameters :
Item | Specifications |
Power supply of this equipment | Three phase AC380V, 50Hz, 4KW |
Power supply for test sample | Single phase AC220V, 16A |
LWGYD-4 flow rate Sensors | Flow range: 0.667-4.167 L/min, pulse output |
Small water pump | Brand: South, model no.: CHL2-40, 380VAC, 50Hz, 0.55kw Protection class: IP55 Insulation class: F |
Waterproof test range | IPX1 test, IPX2 test, IPX1/IPX2 custom |
IPX1-IPX2 | Drip area: 1000*1000mm |
Diameter of drip hole: 0.4mm | |
Distance between each two neighbor holes:20mm | |
Test flow:1~1.5mm/min(IPX1), 3~3.5mm/min(IPX2) | |
Test time | Set according to the test standard (range 1-9999 seconds), the unit is: second |
Water dripping height | 200mm (automatically controlled by the test equipment) |
Water Drainage | Automatically emptying the tank water after the test is completed, should connected with an air source (the 0.6MPa compressed air source should be prepared by user) |
Equipment body material | Industrial aluminum alloy |
Installation method | Wall Mount |
Turntable | 1-5r/ min, can be set |
Control | PLC & 7” touch screen |
Troubleshooting:
No.: | Failure phenomenon | Reason of issue | Treatment |
1 | Turn on the power switch and the touch screen does not light up | External power supply fails | Reconnect external power supply |
Equipment MCB is disconnected | Turn on the equipment MCB | ||
2 | No action when the start button is pressed | The run time is not cleared | Press Count Clear |
The emergency stop switch is not reset | Reset the emergency stop switch | ||
Inverter alarm | Restart after power off | ||
3 | Little or no water from the needles | Needles are clogged | Replace the needle or remove the needle and blow through the needle hole with airflow |
4 | Unstable flow | Air in the pipeline | Press “Dry Drip Box” in the Manual control interface to exhaust the air, or open the vent of the water pump to exhaust the air |